Kestrel® Tune-up & Calibration Services

Basic Test & Calibration Services

All of our measurements are traceable to the National Institute of Standards and Technology, ensuring the highest level of reliability. Our primary Calibration Standards are sent for calibration in accordance with NIST requirements and based on a regular schedule. Only approved laboratories and NIST themselves are used for these calibration services. 

Every Kestrel Pocket Weather Meter is individually calibrated and tested to be within specification at the time of manufacture as follows:

  1. Every wind impeller, including all replacement impellers, is individually wind-tunnel tested against a time of flight ultrasonic anemometer calibrated by NIST.
  2. Every humidity-measuring instrument receives a two-point NIST-traceable calibration and is verified at a third intermediate value against a factory-calibrated hygrometer.
  3. Temperature and pressure are functionally checked and verified against a NIST-traceable calibrated standard or reverified intermediary standard.

For users who must ensure that their Kestrel is operating within specifications, we offer Factory Calibration and Certification services for any Kestrel. For many users, simple recalibration to original specifications, without certification or individual documentation, is sufficient. 

Test Only
Price
Wind Speed / Air Flow Verified in wind tunnel at single speed. Standard and measured values recorded in R/A, together with recommended action (new impeller if out of specification. *Test may also be performed after calibration at customer's request - only one value to be tested/recorded.
$25
Temperature Verified against calibrated standard (calibrated Kestrel 4000 or digital thermometer) at single temperature. Standard and measured values recorded in R/A, together with recommended action (recalibration if out of specification). *Test may also be performed after calibration at customer's request - only one value to be tested/recorded.
$25
Relative Humidity Verified against calibrated standard (calibrated Kestrel 4000 or Edgetech) after equilibration. Standard and measured values recorded in R/A, together with recommended action (recalibration if out of specification). *Test may also be performed after calibration at customer's request - only one value to be tested/recorded.
$25
Barometric Pressure Verified against calibrated standard (calibrated Kestrel 4000 or Mensor) at single pressure. Standard and measured values recorded in R/A, together with recommended action (recalibration if out of specification). *Test may also be performed after calibration at customer's request - only one value to be tested/recorded.
$25
Magnetic Direction Performance verified at four major compass axes. Self-calibration routine performed if needed. Standard and measured values recorded in R/A, together with recommended action.
$25

Take advantage of our Tune-up & Calibration service by filling out a Return Autorization form:


Basic Calibration (with Certificate of Conformity) Price
Wind Speed / Air Flow New calibrated impeller installed (calibrated at two points). Certificate of Wind Speed Conformity issued with SN of unit, Date of Test and operator's initials.
$40
Temperature Standard new unit calibration procedure applied. Certificate of Temperature Conformity issued with SN of unit, Date of Test and operator's initials.
$15
Relative Humidity Standard new unit calibration procedure applied. Certificate of Relative Humidity Conformity issued with SN of unit, Date of Test and operator's initials.
$40
Barometric Pressure Standard new unit calibration procedure applied. Certificate of Pressure Conformity issued with SN of unit, Date of Test and operator's initials.
$25
Magnetic Direction Unit is demagnetized and tested for operation at four points in a known magnetic field through upload and verification of stored values. Self-calibration routine is executed and performance is verified at four major compass axes. Certificate of Direction Conformity issued with SN of unit, Date of Test, and operator's initials.
$25

Take advantage of our Tune-up & Calibration service by filling out a Return Autorization form:


Test & Basic Calibration Price
Wind Speed / Air Flow Perform both test & calibration services (above).
$50
Temperature Perform both test & calibration services (above).
$30
Relative Humidity Perform both test & calibration services (above).
$50
Barometric Pressure Perform both test & calibration services (above).
$40
Magnetic Direction Perform both test & calibration services (above).
$40

Take advantage of our Tune-up & Calibration service by filling out a Return Autorization form: